COBAS 6000 E 601 MODULE E601

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-15 for COBAS 6000 E 601 MODULE E601 manufactured by Roche Diagnostics.

Event Text Entries

[86341542] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[86341543] The customer received a questionable low elecsys total psa immunoassay result for one patient sample. The initial result from an aliquot was 0. 08 ug/l and was reported outside the laboratory. The primary sample was repeated and the result was 6. 5 ug/l. The aliquot was then repeated and the result was 6. 5 ug/l or 6. 4 ug/l. There was no allegation of an adverse event. The reagent lot number was 18561101 with an expiration date of 30-sep-2016.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02007
MDR Report Key6869600
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-15
Date of Report2017-09-28
Date of Event2016-01-06
Date Mfgr Received2016-01-06
Date Added to Maude2017-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeLTJ
Date Received2017-09-15
Model NumberE601
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-09-15
Model NumberE601
Catalog NumberASKU
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.