HYPERBARIC CHAMBER 3600E/ER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a other,user facility report with the FDA on 2017-09-17 for HYPERBARIC CHAMBER 3600E/ER manufactured by Sechrist Industries, Inc..

Event Text Entries

[87317784] Sechrist technician was dispatched onsite and evaluated decompression and emergency vent on the chamber. Technician found that exhaust screen was dirty, removed screen and chamber decompresses normally. Technician instructed customer to remove, inspect screen and clean out at least once month. A device history review was performed (dhr) chamber 3600e/er serial number (b)(4) was manufactured on 03/12/2006. There is no indication that there were any relevant discrepancies during manufacturing. A review of the device history record (dhr) found no non-conformance that could cause or contribute to the reported issue.
Patient Sequence No: 1, Text Type: N, H10


[87317785] Sechrist is submitting this medwatch report in response to medsun report # (b)(4). On (b)(6) 2017 customer reported decompression issues with their hyperbaric chamber 3600e/er. Customer stated that the emergency vent is slow and decompression rate is also slow.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2020676-2017-00023
MDR Report Key6871400
Report SourceOTHER,USER FACILITY
Date Received2017-09-17
Date of Report2017-06-26
Date of Event2017-06-26
Date Mfgr Received2017-06-26
Device Manufacturer Date2006-03-12
Date Added to Maude2017-09-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MAJID MASHAYEKH
Manufacturer Street4225 E. LA PALMA AVENUE
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal92807
Manufacturer G1SECHRIST INDUSTRIES, INC.
Manufacturer Street4225 E. LA PALMA AVENUE
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal Code92807
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHYPERBARIC CHAMBER 3600E/ER
Generic NameHYPERBARIC CHAMBER 3600E/ER
Product CodeCBF
Date Received2017-09-17
Model Number3600E/ER
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSECHRIST INDUSTRIES, INC.
Manufacturer Address4225 E. LA PALMA AVENUE ANAHEIM CA 92807 US 92807