Patient Sequence No: 1, Text Type: N, H10


[86707870] It was reported that the device was overheating before a procedure. No injuries or complications were reported.
Patient Sequence No: 1, Text Type: D, B5


[117305757] Device investigation narrative - the device was received for evaluation. There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issue was observed. Functional evaluation revealed the ballast fan shuts down when unit is powered up, and ballast overheats. The complaint of overheating was confirmed. Cause of fan malfunction is a defective mcu pcb. Unit passes functional testing and fans perform as expected with a known good mcu pcb installed. The complaint investigation has concluded the cause of the failure to be a defective electronic component. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. Device evaluated by the manufacturer. Evaluation codes updated. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1643264-2017-00554
MDR Report Key6878266
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-09-20
Date of Report2017-10-17
Date of Event2017-08-31
Date Mfgr Received2017-10-16
Date Added to Maude2017-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JIM GONZALES
Manufacturer Street7000 W WILLIAM CANNON DR
Manufacturer CityAUSTIN TX 78735
Manufacturer CountryUS
Manufacturer Postal78735
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street76 S. MERIDIAN AVE.
Manufacturer CityOKLAHOMA CITY OK 731076512
Manufacturer CountryUS
Manufacturer Postal Code731076512
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLIGHT SOURCE, 500XL, XENON
Generic NameIMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Product CodeFFS
Date Received2017-09-20
Returned To Mfg2017-09-08
Model Number72200568
Catalog Number72200568
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address76 S. MERIDIAN AVE. OKLAHOMA CITY OK 731076512 US 731076512


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-20