COBAS E-ELECSYS ANTI-CCP 05031656190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-20 for COBAS E-ELECSYS ANTI-CCP 05031656190 manufactured by Roche Diagnostics.

Event Text Entries

[88416806] (b)(4). This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[88416807] The customer received questionable high elecsys anti-ccp immunoassay results for one patient compared to other methodologies. On (b)(6) 2017, the anti-ccp result from a cobas 6000 e 601 module was 163. 6 u/ml. Based on this result, the patient was diagnosed with an autoimmune disease. Sample 1 drawn on (b)(6) 2017 was tested on a quantalyser 160 (elisa) on (b)(6) 2017 and had a result of 3. 35 units. This result was believed to be incorrect and the customer asked the patient to repeat the anti-ccp testing. Sample 2 was drawn on (b)(6) 2017. The result from the quantalyser 160 (elisa) on (b)(6) 2017 was 0. 00 units. An aliquot of the same sample was tested on siemens centaur (clia) on (b)(6) 2017 and the result was <0. 40 u/ml. An aliquot of the same sample was tested on a cobas 6000 e 601 module on (b)(6) 2017 and the result was 146. 40 u/ml. The results were reported outside of the laboratory. There was no allegation of an adverse event. The serial numbers of the involved cobas 6000 e 601 modules were requested but were not provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02022
MDR Report Key6878942
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-20
Date of Report2017-11-06
Date of Event2017-08-25
Date Mfgr Received2017-08-29
Date Added to Maude2017-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS E-ELECSYS ANTI-CCP
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2017-09-20
Model NumberNA
Catalog Number05031656190
Lot Number227465
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-20