ARCHITECT I2000SR ANALYZER 03M74-02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-09-20 for ARCHITECT I2000SR ANALYZER 03M74-02 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[87467594] An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[87467595] He customer questioned an architect total psa result of 1. 218 ng/ml for a patient because it was not compatible with the free psa result of 2. 317 ng/ml. The sample retested at 10. 521 and 10. 383 ng/ml for total psa which matched the repeat free psa results of 2. 052 and 2. 053 ng/ml. The falsely decreased initial total psa result of 1. 218 ng/ml was not reported out of the laboratory. The sample probe was cleaned to resolve the imprecision issue. No adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2017-00348
MDR Report Key6880455
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2017-09-20
Date of Report2017-10-17
Date of Event2017-08-15
Date Mfgr Received2017-10-13
Date Added to Maude2017-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameARCHITECT I2000SR ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeLTJ
Date Received2017-09-20
Catalog Number03M74-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038

Device Sequence Number: 1

Brand NameARCHITECT I2000SR ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeJJE
Date Received2017-09-20
Catalog Number03M74-02
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-20