MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-09-20 for ARCHITECT I2000SR ANALYZER 03M74-02 manufactured by Abbott Manufacturing Inc.
[87467594]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10
[87467595]
He customer questioned an architect total psa result of 1. 218 ng/ml for a patient because it was not compatible with the free psa result of 2. 317 ng/ml. The sample retested at 10. 521 and 10. 383 ng/ml for total psa which matched the repeat free psa results of 2. 052 and 2. 053 ng/ml. The falsely decreased initial total psa result of 1. 218 ng/ml was not reported out of the laboratory. The sample probe was cleaned to resolve the imprecision issue. No adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1628664-2017-00348 |
| MDR Report Key | 6880455 |
| Report Source | FOREIGN,HEALTH PROFESSIONAL |
| Date Received | 2017-09-20 |
| Date of Report | 2017-10-17 |
| Date of Event | 2017-08-15 |
| Date Mfgr Received | 2017-10-13 |
| Date Added to Maude | 2017-09-20 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 0 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Reporter Occupation | MEDICAL TECHNOLOGIST |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
| Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
| Manufacturer City | ABBOTT PARK IL 600643537 |
| Manufacturer Country | US |
| Manufacturer Postal | 600643537 |
| Manufacturer Phone | 224667-512 |
| Manufacturer G1 | ABBOTT MANUFACTURING INC |
| Manufacturer Street | 1921 HURD DRIVE |
| Manufacturer City | IRVING TX 75038 |
| Manufacturer Country | US |
| Manufacturer Postal Code | 75038 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | ARCHITECT I2000SR ANALYZER |
| Generic Name | AUTOMATED IMMUNOASSAY ANALYZER |
| Product Code | LTJ |
| Date Received | 2017-09-20 |
| Catalog Number | 03M74-02 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Eval'ed by Mfgr | N |
| Device Sequence No | 0 |
| Device Event Key | 0 |
| Manufacturer | ABBOTT MANUFACTURING INC |
| Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
| Brand Name | ARCHITECT I2000SR ANALYZER |
| Generic Name | AUTOMATED IMMUNOASSAY ANALYZER |
| Product Code | JJE |
| Date Received | 2017-09-20 |
| Catalog Number | 03M74-02 |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Age | DA |
| Device Eval'ed by Mfgr | N |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ABBOTT MANUFACTURING INC |
| Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2017-09-20 |