MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-09-20 for ARCHITECT I2000SR ANALYZER 03M74-02 manufactured by Abbott Manufacturing Inc.
[87467594]
An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10
[87467595]
He customer questioned an architect total psa result of 1. 218 ng/ml for a patient because it was not compatible with the free psa result of 2. 317 ng/ml. The sample retested at 10. 521 and 10. 383 ng/ml for total psa which matched the repeat free psa results of 2. 052 and 2. 053 ng/ml. The falsely decreased initial total psa result of 1. 218 ng/ml was not reported out of the laboratory. The sample probe was cleaned to resolve the imprecision issue. No adverse impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1628664-2017-00348 |
MDR Report Key | 6880455 |
Report Source | FOREIGN,HEALTH PROFESSIONAL |
Date Received | 2017-09-20 |
Date of Report | 2017-10-17 |
Date of Event | 2017-08-15 |
Date Mfgr Received | 2017-10-13 |
Date Added to Maude | 2017-09-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | MEDICAL TECHNOLOGIST |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NOEMI ROMERO-KONDOS, RN BSN |
Manufacturer Street | 100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3 |
Manufacturer City | ABBOTT PARK IL 600643537 |
Manufacturer Country | US |
Manufacturer Postal | 600643537 |
Manufacturer Phone | 224667-512 |
Manufacturer G1 | ABBOTT MANUFACTURING INC |
Manufacturer Street | 1921 HURD DRIVE |
Manufacturer City | IRVING TX 75038 |
Manufacturer Country | US |
Manufacturer Postal Code | 75038 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARCHITECT I2000SR ANALYZER |
Generic Name | AUTOMATED IMMUNOASSAY ANALYZER |
Product Code | LTJ |
Date Received | 2017-09-20 |
Catalog Number | 03M74-02 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | N |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Brand Name | ARCHITECT I2000SR ANALYZER |
Generic Name | AUTOMATED IMMUNOASSAY ANALYZER |
Product Code | JJE |
Date Received | 2017-09-20 |
Catalog Number | 03M74-02 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 1921 HURD DRIVE IRVING TX 75038 US 75038 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-09-20 |