WARMETTE 56948

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2017-09-21 for WARMETTE 56948 manufactured by Natus Medical Incorporated.

Event Text Entries

[88215271] Both biomedical technicians in contact with tech support reported that there was heat damage apparent on the gasket between plenum (inside which are the heating elements) and blower. When airflow no longer flows to the plenum, heat will build in the plenum and gasket. The customer reported that the thermal cutout was not opened, which cause the heaters to disengage. The customer is instructed to test annually per the user manual 56948-ima-en0408o. A replacement part of the blower motor, a required component was sent due to normal wear and tear on the 11 year old product. Issue was resolved. No further evaluation is needed.
Patient Sequence No: 1, Text Type: N, H10


[88215272] Natus medical received a complaint that on (b)(6) 2017 the warmette had a burning smell. The gasket between the warmette blower and plenum was deformed. Follow-up questionnaire on august 25th, 2017 biomedical technician responded that the "unit developed volumes of smoke inside of unit. " no reported injury or patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2017-00716
MDR Report Key6884104
Report SourceUSER FACILITY
Date Received2017-09-21
Date of Report2017-08-22
Date Mfgr Received2017-08-22
Device Manufacturer Date2006-04-09
Date Added to Maude2017-09-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARC RIVAS
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685142
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWARMETTE
Generic NameWARMETTE
Product CodeLHC
Date Received2017-09-21
Model Number56948
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-21