IOLMASTER 500 000000-1692-983

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-21 for IOLMASTER 500 000000-1692-983 manufactured by Carl Zeiss Meditec Ag (jena).

Event Text Entries

[87188484] The health care professional (hcp) reported that the measurements from the iolmaster 500 were too large. The hcp further reported that he ordered an incorrect implantable collamer lens (icl) based on the iolmaster measurement values and implanted it in the patient's eye. After the icl surgery, the patient suffered from a glaucoma attack. The hcp could not provide any further details about the event or the patient outcome.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615030-2017-00018
MDR Report Key6884828
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-21
Date of Report2017-08-22
Date Mfgr Received2017-10-24
Device Manufacturer Date2015-01-09
Date Added to Maude2017-09-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (JENA)
Manufacturer StreetCARL ZEISS PROMENADE 10
Manufacturer CityJENA, THUERINGIA 07745
Manufacturer CountryGM
Manufacturer Postal Code07745
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameIOLMASTER 500
Generic NameBIOMICROSCOPE, SLIT-LAMP, AC-POWERED
Product CodeHJO
Date Received2017-09-21
Returned To Mfg2017-09-27
Model NumberNA
Catalog Number000000-1692-983
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (JENA)
Manufacturer AddressCARL ZEISS PROMENADE 10 JENA, THUERINGIA 07745 GM 07745


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-21