ROHO? HIGH PROFILE? SINGLE COMPARTMENT CUSHION 1R1112C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer report with the FDA on 2017-09-24 for ROHO? HIGH PROFILE? SINGLE COMPARTMENT CUSHION 1R1112C manufactured by Roho, Inc..

Event Text Entries

[87179792] The patient stated he does not check cushion inflation per the safety information in the cushion operation manual and labels. Also, while the customer claimed an injury, roho has not seen medical records confirming this. The cushion was evaluated, but roho was unable to determine if there was a manufacturing defect due to the patches put in place by the customer. The cushion was also being used two years past the warranty period.
Patient Sequence No: 1, Text Type: N, H10


[87179793] Patient contacted roho and stated that his cushion went flat and it caused a pressure sore that required hospitalization and surgery. The cushion started to go down and he did not realize it because he doesn't routinely perform hand checks.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1419507-2017-00005
MDR Report Key6889224
Report SourceCONSUMER
Date Received2017-09-24
Date of Report2017-09-21
Date of Event2017-04-01
Date Mfgr Received2017-08-24
Device Manufacturer Date2013-05-28
Date Added to Maude2017-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJOHN KECK
Manufacturer Street100 N FLORIDA
Manufacturer CityBELLEVILLE IL 62221
Manufacturer CountryUS
Manufacturer Postal62221
Manufacturer Phone8008513449
Manufacturer G1ROHO, INC.
Manufacturer Street1501 S. 7TH ST.
Manufacturer CityBELLEVILLE IL 62223
Manufacturer CountryUS
Manufacturer Postal Code62223
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameROHO? HIGH PROFILE? SINGLE COMPARTMENT CUSHION
Generic NameCUSHION, FLOTATION, PRODUCT CODE KIC
Product CodeKIC
Date Received2017-09-24
Returned To Mfg2017-09-21
Model Number1R1112C
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerROHO, INC.
Manufacturer Address100 N. FLORIDA AVE. BELLEVILLE IL 62221 US 62221


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2017-09-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.