MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-25 for COBAS E 411 IMMUNOASSAY ANALYZER E411 RACK 04775201001 manufactured by Roche Diagnostics.
[89572401]
This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10
[89572402]
The customer stated that they received an erroneous result for one patient sample tested for the elecsys vitamin b12 immunoassay (b12) on a cobas e 411 immunoassay analyzer (e411). The erroneous result was not reported outside of the laboratory. The sample initially resulted as 808 pg/ml. The sample was repeated twice, resulting as 408 pg/ml and 398 pg/ml. The result of 408 ng/ml was considered to be correct and was reported outside of the laboratory. No adverse events were alleged to have occurred with the patient. The b12 reagent lot number was 241742-02, with an expiration date of (b)(6) 2018. The field service engineer ran performance testing on the analyzer. No further issues have been reported by the customer.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 1823260-2017-02061 |
MDR Report Key | 6889512 |
Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
Date Received | 2017-09-25 |
Date of Report | 2017-10-09 |
Date of Event | 2017-09-01 |
Date Mfgr Received | 2017-09-01 |
Date Added to Maude | 2017-09-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 0 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | NA MICHAEL LESLIE |
Manufacturer Street | 9115 HAGUE ROAD NA |
Manufacturer City | INDIANAPOLIS IN 46250 |
Manufacturer Country | US |
Manufacturer Postal | 46250 |
Manufacturer Phone | 3175214343 |
Manufacturer G1 | HITACHI HIGH TECH CORP. |
Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
Manufacturer City | IBARAKI 312-8504 |
Manufacturer Country | JA |
Manufacturer Postal Code | 312-8504 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | COBAS E 411 IMMUNOASSAY ANALYZER |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | CDD |
Date Received | 2017-09-25 |
Model Number | E411 RACK |
Catalog Number | 04775201001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | R |
Device Sequence No | 0 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Brand Name | COBAS E 411 IMMUNOASSAY ANALYZER |
Generic Name | IMMUNOCHEMISTRY ANALYZER |
Product Code | JJE |
Date Received | 2017-09-25 |
Model Number | E411 RACK |
Catalog Number | 04775201001 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ROCHE DIAGNOSTICS |
Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-09-25 |