TUR IRRIGATION SET ARC4005P

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-09-26 for TUR IRRIGATION SET ARC4005P manufactured by Baxter Healthcare - Cali.

Event Text Entries

[87316086] Initial reporter phone: (b)(6). The device has been received and the evaluation is in progress. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[87316087] It was reported that a tur irrigation set had a loose connection between the tubing and the connector. This occurred before use. There was no patient involvement. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


[104200783] The device was received for evaluation. Visual inspection revealed that the rubber tube was separated from the guide tube. The reported condition was verified. The cause of the condition was determined to be an assembly issue during manufacturing. A nonconformance has been opened to address this issue. A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot. Should additional relevant information become available, a supplemental report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[104200784]
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1416980-2017-07824
MDR Report Key6893260
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-09-26
Date of Report2017-11-01
Date of Event2017-08-28
Date Mfgr Received2017-10-20
Device Manufacturer Date2017-07-20
Date Added to Maude2017-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone2242702068
Manufacturer G1BAXTER HEALTHCARE - CALI
Manufacturer StreetCALLE 36 NO. 2C-22 APARTADO AERO 2446
Manufacturer CityCALI
Manufacturer CountryCO
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTUR IRRIGATION SET
Generic NameSYSTEM, IRRIGATION, UROLOGICAL
Product CodeLJH
Date Received2017-09-26
Returned To Mfg2017-09-08
Model NumberNA
Catalog NumberARC4005P
Lot NumberSE17GI3
ID NumberNA
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE - CALI
Manufacturer AddressCALI


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.