EXPEL? MULTO-PURPOSE DRAINAGE CATHETER 39311-105

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-26 for EXPEL? MULTO-PURPOSE DRAINAGE CATHETER 39311-105 manufactured by Boston Scientific Corporation.

Event Text Entries

[87385021]
Patient Sequence No: 1, Text Type: N, H10


[87385022] Patient returned to hospital for dressing change due to leaking abscess drain placed in ir ct. The tubing and stopcock were not the source of the leak. After rt evaluated tube the leak was from the catheter hub being cracked. Patient was kept and pre procedure work done in spec procedure holding area. Dr. Ordered for drainage catheter exchange and tube was exchanged under x-ray guidance. Patient was recovered post procedure.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6893949
MDR Report Key6893949
Date Received2017-09-26
Date of Report2017-08-31
Date of Event2017-08-03
Report Date2017-08-31
Date Reported to FDA2017-08-31
Date Reported to Mfgr2017-08-31
Date Added to Maude2017-09-26
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXPEL? MULTO-PURPOSE DRAINAGE CATHETER
Generic NameDRAINAGE CATHETER
Product CodeOEI
Date Received2017-09-26
Returned To Mfg2017-08-08
Model Number39311-105
Lot Number20546327
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC CORPORATION
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.