CLINITRON AT*HOME P0800010016

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-27 for CLINITRON AT*HOME P0800010016 manufactured by Hill-rom Batesville.

Event Text Entries

[87586817] Hill-rom confirmed the patient remained on the bed while it was not working and that the patients family placed a non hill-rom air mattress on top a few days after the malfunction. The account stated that the patients health has been declining for some time now, due to his over all diagnosis. Development of pressure ulcers is multifactorial and cannot be only attributed to performance of the surface. Risk factors include protein-calorie malnutrition, microclimate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy. Position changes are key to pressure sore prevention and treatment. These changes need to be frequent, repositioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas. The hill-rom technician found the blower, diffuser board, microspheres and air cushions needed to be replaced. The technician replaced the blower, diffuser board, microspheres and air cushions to resolve the issue. Based on this information, no further action is required. Per the hill-rom user manual, poor fluidization; if fluidization is sluggish or uneven, notify your hill-rom representative. Fluidization is affected by the following: room temperature, humidity, the amount of materials, such as fluid, cells or cellular debris, which has escaped from the blood vessels and has been deposited in tissues or on tissue surfaces, restricted air circulation from blankets on the bed. The wear of microspheres to the point that they will not fluidize properly is multifaceted and their replacement is a practical occurrence that is inherent with use of the bed. Several factors come into account, environment, use and care of the patient. Hill-rom recommends not exceeding an ambient room temperature of 75 degrees f. The microspheres can handle limited amounts of fluids passing through the filter sheet. Excessive humidity, incontinence and bodily fluids will saturate the beads and hamper fluidization. Treatment from caregivers such as petroleum based topical ointments and silver compounds will ruin the coating on the beads and permanently destroy their fluidizing properties. A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2013. It is unknown if the facility performed any other preventative maintenance on this bed.
Patient Sequence No: 1, Text Type: N, H10


[87586818] Hill-rom received a report from the account stating the patient developed a new stage 4 pressure ulcer on his buttocks. The bed was located at the account. This report was filed in our complaint handling system as complaint #(b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1824206-2017-00400
MDR Report Key6898139
Date Received2017-09-27
Date of Report2017-09-15
Date of Event2017-09-15
Date Mfgr Received2017-09-15
Date Added to Maude2017-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT FAMILY MEMBER OR FRIEND
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTONY WERNER
Manufacturer Street1069 STATE ROUTE 46 EAST
Manufacturer CityBATESVILLE IN 47006
Manufacturer CountryUS
Manufacturer Postal47006
Manufacturer Phone8129312359
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameCLINITRON AT*HOME
Generic NameBED, AIR FLUIDIZED
Product CodeINX
Date Received2017-09-27
Model NumberP0800010016
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHILL-ROM BATESVILLE
Manufacturer Address1069 STATE ROUTE 46 EAST BATESVILLE IN 47006 US 47006


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-27