THERASEED PD-103 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-27 for THERASEED PD-103 NA manufactured by Theragenics Corporation.

Event Text Entries

[87858180] Upon review of the order form, it was confirmed that the units of activity ordered by the customer was entered incorrectly during the order entry process. The customer requested 1. 7 mci and the order was entered as 1. 7 u. The customer service representative did not confirm the order activity with the customer prior to entry. The activity was entered lower than requested by the customer. The error was not detected during the order entry verification process prior to shipment of the order nor during the customer's order confirmation process. Upon quality review of the received order at the users facility, the discrepancy was detected, and the planned treatment was cancelled. There was no failure or malfunction. The seed order was filled with a lower than requested activity, which would have resulted in a misadministration.
Patient Sequence No: 1, Text Type: N, H10


[87858181] Seed activity for brachytherapy order was entered using the wrong units causing a discrepancy between the ordered seed strength and the seed strength that was delivered. The error was detected at the user facility prior to use and the treatment plan and order was cancelled.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1037598-2017-00001
MDR Report Key6898273
Date Received2017-09-27
Date of Report2017-04-25
Date of Event2017-04-19
Report Date2017-04-25
Date Reported to Mfgr2017-04-25
Date Mfgr Received2017-04-25
Device Manufacturer Date2017-04-15
Date Added to Maude2017-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMS MELISA GILCHRIST
Manufacturer Street5203 BRISTOL INDUSTRIAL WAY
Manufacturer CityBUFORD GA 30518
Manufacturer CountryUS
Manufacturer Postal30518
Manufacturer Phone7702710233
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTHERASEED PD-103
Generic NameRADIONUCLIDE BRACHYTHERPY SOURCE
Product CodeIWH
Date Received2017-09-27
Model NumberPD-103
Catalog NumberNA
Lot Number1714J
Device Expiration Date2017-05-12
OperatorOTHER HEALTH CARE PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTHERAGENICS CORPORATION
Manufacturer Address5203 BRISTOL INDUSTRIAL WAY BUFORD GA 30518 US 30518


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-27