PATHROMTIN SL 10446067

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-09-29 for PATHROMTIN SL 10446067 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[89422537] Siemens healthcare diagnostics reviewed the backup files provided by the customer of the sysmex cs-5100 analyzer to determine the cause of the high activated partial prothrombin time (aptt) result. Review of the sysmex cs-5100 analyzer backup file included analysis of the cs-5100 kinetics, which showed no issues and no error codes were triggered. The system operated as specified and a non-conformance could not be identified. The instrument is performing according to specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


[89422538] A discordant, high activated partial thromboplastin time (aptt) result of 115 seconds on one patient sample was generated on the sysmex cs-5100 analyzer using the pathromtin sl assay (lot 536693). This result was not reported to the physician. The same sample, using a different lot of pathromtin sl reagent (lot 536695), was tested on a bcs xp system generating an aptt result of 35. 5 seconds. This result was in line with the clinical picture of the patient. There are no reports of patient intervention or adverse health consequence due to the high aptt result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00105
MDR Report Key6904643
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-09-29
Date of Report2017-11-15
Date of Event2017-07-10
Date Mfgr Received2017-11-15
Device Manufacturer Date2016-10-06
Date Added to Maude2017-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLORIANN RUSSO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242287
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePATHROMTIN SL
Generic NamePATHROMTIN SL
Product CodeGFO
Date Received2017-09-29
Catalog Number10446067
Lot Number536693A
Device Expiration Date2018-05-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-09-29