HARVEST TERUMO 51401

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-09-29 for HARVEST TERUMO 51401 manufactured by Terumo Bct.

Event Text Entries

[89529040] Investigation: an unused harvest terumo platelet concentrate procedure pack was returnedfor investigation. Upon visual inspection, it was noted that the expiration date on the outsitde label of the set is confirmed to read 2017-08. The expiration date of the label on the lower packaging tray (plasma products pack) is confirmed to read 2016-07. Ac expiration date isconfirmed to read 2016-07. Harvest product set contain multiple components that have various expiry dates. The outer kitlabel contains the expiry date associated with the component that has the shortest expirytimeframe. Investigation is in process. A follow-up report will be provided.
Patient Sequence No: 1, Text Type: N, H10


[89529041] The customer reported that after opening a harvest terumo platelet concentrate procedure pack, it was noticed that the expiry date on the label of a bottle of anti-coagulant (ac) was 07/2016. The pack's outer set was labeled with an expiry date of 08/2017. No patient (donor) was connected at the time of the event. No patient (donor) information isreasonably known.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1722028-2017-00383
MDR Report Key6906198
Date Received2017-09-29
Date of Report2017-09-29
Date of Event2017-09-07
Date Mfgr Received2018-05-14
Device Manufacturer Date2014-09-26
Date Added to Maude2017-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVE KERN
Manufacturer Street10810 W. COLLINS AVE
Manufacturer CityLAKEWOOD CO 80215
Manufacturer CountryUS
Manufacturer Postal80215
Manufacturer Phone3032392246
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHARVEST TERUMO
Generic NamePC-60 PLATELET CONC PRODECURE PACK
Product CodeORG
Date Received2017-09-29
Returned To Mfg2017-09-16
Catalog Number51401
Lot Number503985
ID Number05020583514010
Device Expiration Date2016-07-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTERUMO BCT
Manufacturer AddressLAKEWOOD CO 80215 US 80215


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-09-29

© 2026 FDA.report