[88053464]
Patient Sequence No: 1, Text Type: N, H10


[88053465] Stryker is using a very poor quality dvi connector for their video integration system. This type of connector keeps breaking very often, almost weekly. They are using a dvi connector that is not intended to be used for the amount of mechanical work that it is used for. We have 13 operating rooms in our facility and we replace this connector almost weekly. They break because they cannot withstand the weight of the cables that are used and also the amount of times that the dvi connector needs to be used on the daily basis. Manufacturer response for video integration system, (per site reporter): they have replaced the connector multiple times, but haven't come up with a solution.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6910415
MDR Report Key6910415
Date Received2017-10-03
Date of Report2017-09-29
Date of Event2016-06-01
Report Date2017-09-29
Date Reported to FDA2017-09-29
Date Reported to Mfgr2017-09-29
Date Added to Maude2017-10-03
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSWITCHPOINT INFINITY
Generic NameENDOSCOPIC VIDEO IMAGING SYSTEM/COMPONENT, GASTROENTEROLOGY-UROLOGY
Product CodeOCS
Date Received2017-10-03
Model Number0678001000
Catalog Number0678-001-000
Device AvailabilityY
Device Age2 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER COMMUNICATIONS INC.
Manufacturer Address571 SILVERON BLVD. FLOWER MOUND TX 75028 US 75028


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-03

© 2026 FDA.report