FUSION COMPACT NAVIGATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-10-03 for FUSION COMPACT NAVIGATION SYSTEM manufactured by Medtronic Navigation, Inc..

Event Text Entries

[88060853]
Patient Sequence No: 1, Text Type: N, H10


[88060854] A female patient was scheduled to undergo a bilateral functional endoscopic sinus surgery (fess) procedure. She was brought to the or and induced with general anesthesia. The navigation system was brought onto the field and it initially malfunctioned. The surgeon was able to successful register the patient's anatomy to the pre-operative ct scan, but was unable to register the instruments (suctions) to the system. The team was able to trouble shoot the problem with help from the company via phone and were able to get the equipment to properly register. The company rep suggested that the staff post trouble-shooting instructions (4 specific actions) in the room.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6910565
MDR Report Key6910565
Date Received2017-10-03
Date of Report2017-09-28
Date of Event2017-09-07
Report Date2017-09-28
Date Reported to FDA2017-09-28
Date Reported to Mfgr2017-09-28
Date Added to Maude2017-10-03
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFUSION COMPACT NAVIGATION SYSTEM
Generic NameEAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Product CodeOGR
Date Received2017-10-03
OperatorPHYSICIAN
Device AvailabilityN
Device Age7 YR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC NAVIGATION, INC.
Manufacturer Address826 COAL CREEK CIRCLE LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-03