COBAS 6000 E 601 MODULE E601 04745922001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-03 for COBAS 6000 E 601 MODULE E601 04745922001 manufactured by Roche Diagnostics.

Event Text Entries

[89620191] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[89620192] The customer complained of discrepant elecsys brahms pct (pct) result for one patient sample. The initial pct result was 0. 110 ng/ml. This result was reported outside of the laboratory. The initial result was questioned so a repeat test was ordered. The repeat pct result was >100 ng/ml, with a data flag. Repeat testing was performed on another analyzer and deemed to be correct. There was no adverse event from the initial result being reported out. Qc results prior to the event were acceptable. The pct reagent lot was 24279802 with an expiration date of 30-sep-2018. The field engineering specialist was not able to reproduce the error or find a root cause. He checked the fluidics and mechanical adjustments. He performed precision and performance testing which all passed. Further investigation could not determine a root cause. A historical query for this site was performed and no new or past escalated complaints of this nature on any like instruments were found for the past 12 months.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02184
MDR Report Key6911540
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-10-03
Date of Report2017-12-11
Date of Event2017-09-15
Date Mfgr Received2017-09-15
Date Added to Maude2017-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodePMT
Date Received2017-10-03
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457

Device Sequence Number: 1

Brand NameCOBAS 6000 E 601 MODULE
Generic NameIMMUNOCHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-10-03
Model NumberE601
Catalog Number04745922001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.