ARCHITECT C16000 SYSTEM 03L77-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-10-03 for ARCHITECT C16000 SYSTEM 03L77-01 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[89630604] Local phone number continued: 912699 an abbott field service engineer (fse) completed troubleshooting of the instrument. This involved removal and cleaning of waste tubing, water shock treatment to the waterbath and sample and syringe tubing were evaluated and were deemed ok. Controls and precision testing was completed and the instrument was put back into service. Review of the instrument log review was completed; from (b)(6) to (b)(6) 2017, there were 146 occurrences of error code 3375 (unable to process test, aspiration error occurred. ) tracking and trending did not identify an adverse trend for the customer issue. Labeling was reviewed and found to be adequate. Based on the available information no product deficiency of the architect c16000 analyzer, list number 03l77, was identified.
Patient Sequence No: 1, Text Type: N, H10


[89630605] The customer generated a falsely elevated potassium result while using the architect c16000 analyzer. The customer indicated sid 0683985h generated an initial result of 6. 8 and a retest result of 4. 9 mmol/l there was no adverse impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1628664-2017-00368
MDR Report Key6911988
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2017-10-03
Date of Report2018-09-05
Date of Event2017-09-10
Date Mfgr Received2018-08-20
Device Manufacturer Date2009-05-01
Date Added to Maude2017-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Removal Correction Number1628664-07/17/18-001-C
Event Type3
Type of Report3

Device Sequence Number: 0

Brand NameARCHITECT C16000 SYSTEM
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeCEM
Date Received2017-10-03
Catalog Number03L77-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038

Device Sequence Number: 1

Brand NameARCHITECT C16000 SYSTEM
Generic NameAUTOMATED CHEMISTRY ANALYZER
Product CodeJJE
Date Received2017-10-03
Catalog Number03L77-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.