SYSMEX SA-10 CR690716

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-10-03 for SYSMEX SA-10 CR690716 manufactured by Sysmex Ra Co. Ltd..

Event Text Entries

[88154309] Investigation by sysmex corporation in (b)(4) determined excessive voltage that exceeded the product standard was applied to the power supply unit. The surge absorber of the board was damaged. There was no abnormality found in the manufacturing of the product. The board was replaced and the instrument was operational. Any incident with excessive heat has the potential to cause user harm. The pcb is made of flame resistant materials and covered with a metal shield, but excessive heat from the event poses a risk of inhalation of smoke and/or harmful vapors from melted materials. The pcb is not accessible to the user, and is grounded, reducing the potential for harm due to shock.
Patient Sequence No: 1, Text Type: N, H10


[88154310] The user reported that after a power interruption, the sampler would not turn back on. There was no visible fire, smoke or odor noted by the user.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000515253-2017-00035
MDR Report Key6912896
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2017-10-03
Date of Report2017-10-03
Date of Event2017-04-19
Date Mfgr Received2017-04-19
Device Manufacturer Date2016-07-07
Date Added to Maude2017-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PETER SHEARSTONE
Manufacturer Street1850-3 HIROOKANOMURA
Manufacturer CitySHIOJIRI, NAGANO IL 399-0702
Manufacturer CountryJA
Manufacturer Postal399-0702
Manufacturer Phone2245439514
Manufacturer G1SYSMEX RA CO. LTD.
Manufacturer Street1850-3 HIROOKANOMURA
Manufacturer CitySHIOJIRI, NAGANO 399-0702
Manufacturer CountryJA
Manufacturer Postal Code399-0702
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYSMEX SA-10
Generic NameSA-10 SAMPLER UNIT
Product CodeJRO
Date Received2017-10-03
Model NumberSA-10
Catalog NumberCR690716
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYSMEX RA CO. LTD.
Manufacturer Address1850-3 HIROOKA-NOMURA SHIOJIRI-SHI, NAGANO 399-0702 JA 399-0702


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.