MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-09 for COBAS 6000 C (501) MODULE C501 04745914001 manufactured by Roche Diagnostics.
[89951188]
(b)(4). The event occurred in: (b)(6).
Patient Sequence No: 1, Text Type: N, H10
[89951189]
The customer complained of a high result for albt2 tina-quant albumin gen. 2. The initial albumin result was 50 mg/l. The sample was repeated with results of 20 mg/l and 20 mg/l. There was no allegation of an adverse event. The erroneous result was not reported outside of the laboratory. The albt2 tina-quant albumin gen. 2 reagent lot was 253367 with an expiration date that was not provided. The pressure on the gear pump was adjusted and a field service representative was dispatched to replace the gear pump head.
Patient Sequence No: 1, Text Type: D, B5
| Report Number | 1823260-2017-02229 |
| MDR Report Key | 6928935 |
| Report Source | FOREIGN,HEALTH PROFESSIONAL,U |
| Date Received | 2017-10-09 |
| Date of Report | 2017-11-01 |
| Date of Event | 2017-09-20 |
| Date Mfgr Received | 2017-09-20 |
| Date Added to Maude | 2017-10-09 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 0 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 3 |
| Event Location | 3 |
| Manufacturer Contact | NA MICHAEL LESLIE |
| Manufacturer Street | 9115 HAGUE ROAD NA |
| Manufacturer City | INDIANAPOLIS IN 46250 |
| Manufacturer Country | US |
| Manufacturer Postal | 46250 |
| Manufacturer Phone | 3175214343 |
| Manufacturer G1 | HITACHI HIGH TECH CORP. |
| Manufacturer Street | 882 ICHIGE HITACHINAKA NA |
| Manufacturer City | IBARAKI 312-8504 |
| Manufacturer Country | JA |
| Manufacturer Postal Code | 312-8504 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Removal Correction Number | NA |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | COBAS 6000 C (501) MODULE |
| Generic Name | CLINICAL CHEMISTRY ANALYZER |
| Product Code | CJW |
| Date Received | 2017-10-09 |
| Model Number | C501 |
| Catalog Number | 04745914001 |
| Lot Number | NA |
| ID Number | NA |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 0 |
| Device Event Key | 0 |
| Manufacturer | ROCHE DIAGNOSTICS |
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
| Brand Name | COBAS 6000 C (501) MODULE |
| Generic Name | CLINICAL CHEMISTRY ANALYZER |
| Product Code | JJE |
| Date Received | 2017-10-09 |
| Model Number | C501 |
| Catalog Number | 04745914001 |
| Lot Number | NA |
| ID Number | NA |
| Operator | HEALTH PROFESSIONAL |
| Device Availability | Y |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ROCHE DIAGNOSTICS |
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2017-10-09 |