MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2017-10-09 for I-STAT ACT KAOLIN CARTRIDGE 03P87-25 manufactured by Abbott Point Of Care.
[90313535]
Apoc incident # (b)(4). Apoc labeling was evaluated during the investigation as pertaining to the event. The investigation was completed on 11/09/2015. Retain and return product was tested and found to be functioning according to specification. Investigation: the device history record for this lot was reviewed. The lot passed finished goods release criteria. Forty retained and twenty-two returned cartridges were tested using blood samples which simulate patients undergoing oral anticoagulant therapy. Testing met the acceptance criteria found in q04. 01. 003 rev. W, appendix 1 - product complaint level 2 and level 3 investigation procedure. No deficiency has been identified. Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat cartridge lot is meeting specification.
Patient Sequence No: 1, Text Type: N, H10
[90313536]
On (b)(6) 2015, abbott point of care was contacted by a customer regarding i-stat act kaolin cartridges that yielded unexpected results on a patient. There was no additional patient information available at the time of this report. Actk results: >1000 (initial test). >1000 (retest 1). 602 (retest 2). At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc; however, this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care (b)(4).
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2245578-2017-00373 |
MDR Report Key | 6932306 |
Report Source | HEALTH PROFESSIONAL |
Date Received | 2017-10-09 |
Date of Report | 2017-10-09 |
Date of Event | 2015-09-23 |
Date Mfgr Received | 2017-09-13 |
Device Manufacturer Date | 2015-05-04 |
Date Added to Maude | 2017-10-09 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | LINDA MACZUSZENKO |
Manufacturer Street | 400 COLLEGE ROAD |
Manufacturer City | PRINCETON NJ 08540 |
Manufacturer Country | US |
Manufacturer Postal | 08540 |
Manufacturer Phone | 6136885949 |
Manufacturer G1 | ABBOTT POINT OF CARE CANADA LTD. |
Manufacturer Street | 185 CORKSTOWN ROAD |
Manufacturer City | OTTAWA, ONTARIO K2H 8V4 |
Manufacturer Country | CA |
Manufacturer Postal Code | K2H 8V4 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | I-STAT ACT KAOLIN CARTRIDGE |
Generic Name | ACT KAOLIN CARTRIDGE |
Product Code | JBP |
Date Received | 2017-10-09 |
Returned To Mfg | 2015-10-05 |
Model Number | NA |
Catalog Number | 03P87-25 |
Lot Number | T15124 |
Device Expiration Date | 2015-10-28 |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT POINT OF CARE |
Manufacturer Address | 400 COLLEGE ROAD PRINCETON NJ 085406607 US 085406607 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-10-09 |