BLADE CLIPPER SURG 4406

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-10-10 for BLADE CLIPPER SURG 4406 manufactured by Carefusion, Inc.

Event Text Entries

[90312108] (b)(4). - the adverse event occurred while the nurse was attempting to attach the blade to the handle stating that the blade is very hard to attach to the clipper because the plastic around the blade are very tiny. (b)(4) examined the handle and found this was caused by physical damage. The edge of driving shaft and? The parts for mounting the blade? Were broken by? Dropping to the hard floor. (b)(4) has never changed plastic material(pom and asa) of the clipper handle. Should further information be received, a follow-up emdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[90312109] A nurse did cut her finger when she had to place the blade on the clipper. The blade is very sharp on the inside and is not protected at all, this is where the accident occurred. (in comparison the neuro blades and sensi blades are both protected at the inner side ). In addition, they mentioned that the blade is very hard to attach to the clipper because the plastic around the blade are very tiny.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423507-2017-00103
MDR Report Key6937659
Date Received2017-10-10
Date of Report2017-10-10
Date of Event2017-09-12
Date Mfgr Received2017-09-12
Date Added to Maude2017-10-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer G1CAREFUSION, INC
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal Code60061
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBLADE CLIPPER SURG
Generic NameRAZOR, SURGICAL
Product CodeLWK
Date Received2017-10-10
Returned To Mfg2017-10-05
Model Number4406
Lot Number0816
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.