TRIAGE D-DIMER PANEL 98100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a consumer,health professional, report with the FDA on 2017-10-11 for TRIAGE D-DIMER PANEL 98100 manufactured by Alere San Diego, Inc..

Event Text Entries

[89236409] Investigation pending.
Patient Sequence No: 1, Text Type: N, H10


[89236410] On (b)(6) 2017, the patient arrived at the physician's office complaining of chest pain. Both d-dimer and tni testing was conducted on the patient. A whole-blood edta sample was collected at 4:30 pm and the triage d-dimer panel produced an abnormal d-dimer result of 1030 ng/ml d-du. The tni results are not available per customer but produced a normal result. Based on the abnormal d-dimer result, the patient was sent to the emergency room for a chest ct. The chest ct showed no evidence of pulmonary embolism and no evidence of acute cardio pulmonary disease. At the emergency room, confirmatory testing was conducted using the hemosil system and resulted in a normal d-dimer value of 300 ng/ml d-du. On (b)(6) 2017, the same whole-blood edta sample was tested an additional two times at approximately 1:00 pm. The triage d-dimer panel produced abnormal results of 968 ng/ml d-du and 825 ng/ml d-du. The facility's normal d-dimer cutoff for both the triage and hemosil systems was reported to be 400 ng/ml d-du.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2017-00153
MDR Report Key6942683
Report SourceCONSUMER,HEALTH PROFESSIONAL,
Date Received2017-10-11
Date of Report2017-09-13
Date of Event2017-09-12
Date Mfgr Received2017-12-01
Date Added to Maude2017-10-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE D-DIMER PANEL
Generic NameFIBRIN SPLIT PRODUCTS
Product CodeGHH
Date Received2017-10-11
Returned To Mfg2017-11-29
Model Number98100
Lot NumberW62923B
Device Expiration Date2018-01-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALERE SAN DIEGO, INC.
Manufacturer Address9975 SUMMERS RIDGE ROAD SAN DIEGO CA 92121 US 92121


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-10-11