SYNGO LAB DATA MANAGER APPLICATION SOFTWARE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-12 for SYNGO LAB DATA MANAGER APPLICATION SOFTWARE manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[90295503] The customer contacted the siemens customer care center (ccc). Ccc connected remotely and reviewed the data logs. The patient sample was placed on the track at 10:30 and the results were in syngo by 10:46 for all tests. The customer has the magnesium set up to hold for magnesium therapeutic which means the technician has to release this test. Review of the audit trail indicated that the magnesium was in held status and was not released until 11:46 by the technician. Syngo performed as intended. Ccc discussed the user error with the laboratory supervisor. The cause of the delay was due to a user error. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[90295504] The customer reported that a magnesium therapeutic patient result took over an hour to be reported from the syngo lab data manager to the laboratory information system. The customer received a complaint from the physician regarding the delay. There are no known reports of patient intervention or adverse health consequences due to the delay.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00773
MDR Report Key6946370
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-10-12
Date of Report2017-10-12
Date of Event2017-09-14
Date Mfgr Received2017-09-15
Date Added to Maude2017-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Generic NameSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Product CodeNSX
Date Received2017-10-12
Model NumberSYNGO LAB DATA MANAGER APPLICATION SOFTWARE
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE NEWARK DE 19702 US 19702


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-12