ORS-100 FLUID WARMER DRAPE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2017-10-12 for ORS-100 FLUID WARMER DRAPE manufactured by Microtek Dominicana, S.a..

Event Text Entries

[90307834] A sample was received for evaluation. The sample drape had a small burn mark/hole about three inches from the center of the drape. The dhr was reviewed for lot d171141 and it was noted that this lot had (b)(4) pcs that were manufactured on 04/24/2017. No defects were reported during quality inspections. Based on the dhr and sample review this issue does not appear to be the result of a personnel, process, or material issue. Melts, though uncommon, can occur when the warmer is not draped properly, not turned off by using the power button, and/or there is insufficient fluid in the warmer basin, contrary to the operations manual, product labeling, product insert and in-service presentations. Because this appears to be related to misuse by the customer no actions are being taken at this time.
Patient Sequence No: 1, Text Type: N, H10


[90307835] End user reported that water was found in the basin of an ors irrigation fluid warmer following the completion of an ortho case. End user stated that no sharps were introduced into the drape/basin during the case. No patient injury or treatment were reported for the incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8043817-2017-00028
MDR Report Key6946450
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2017-10-12
Date of Report2017-09-15
Date of Event2017-09-11
Date Mfgr Received2017-09-15
Date Added to Maude2017-10-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS JENNIFER MRKVICKA
Manufacturer Street1 ECOLAB PLACE
Manufacturer CityST. PAUL MN 55102
Manufacturer CountryUS
Manufacturer Postal55102
Manufacturer Phone6512502237
Manufacturer G1MICROTEK DOMINICANA
Manufacturer StreetZONA FRANCA NO. 2
Manufacturer CityLA ROMANA,
Manufacturer CountryDR
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORS-100 FLUID WARMER DRAPE
Generic NameFLUID WARMING AND SLUSH DRAPES
Product CodeLHC
Date Received2017-10-12
Returned To Mfg2017-10-02
Model NumberORS-100
Lot NumberD171141
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROTEK DOMINICANA, S.A.
Manufacturer AddressZONA FRANCA NO. 2 LA ROMANA, DR


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-12

© 2026 FDA.report