IOLMASTER XP 000000-1322-734

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-13 for IOLMASTER XP 000000-1322-734 manufactured by Carl Zeiss Meditec Ag (jena).

Event Text Entries

[89347686] A number of factors not related to the lolmaster may have influenced the surgical outcome. The quality of keratometry measurements is highly dependent on stable tear film, correct fixation, and wide open eyes. The user manual describes in detail how to perform intraocular lens measurements and calculations and contains warnings about relevant parameters.
Patient Sequence No: 1, Text Type: N, H10


[89347687] The healthcare professional (hcp) reported the following: the iolmaster was used for the patient's biometry measurements. The alcon sa60at lens with a power of +17. 0 diopter was used. The os post-refractive outcome after a cataract surgery with an intraocular lens (iol) implantation differed -6. 00 diopter from the target refraction. The hcp made a decision to exchange the iol.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615030-2017-00022
MDR Report Key6950226
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-10-13
Date of Report2017-09-14
Date of Event2017-08-11
Date Mfgr Received2017-10-18
Device Manufacturer Date2009-09-10
Date Added to Maude2017-10-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (JENA)
Manufacturer StreetCARL ZEISS PROMENADE 10
Manufacturer CityJENA, THUERINGIA 07745
Manufacturer CountryGM
Manufacturer Postal Code07745
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIOLMASTER
Generic NameBIOMICROSCOPE, SLIT-LAMP, AC-POWERED
Product CodeHJO
Date Received2017-10-13
Model NumberXP
Catalog Number000000-1322-734
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (JENA)
Manufacturer AddressCARL ZEISS PROMENADE 10 JENA, THUERINGIA 07745 GM 07745


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-10-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.