CHROMID? MRSA SMART AGAR 413050

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-16 for CHROMID? MRSA SMART AGAR 413050 manufactured by Biom?rieux Sa.

Event Text Entries

[90544932] Device not returned to manufacturer
Patient Sequence No: 1, Text Type: N, H10


[90544934] A customer in (b)(6) reported the occurrence of a (b)(6) result in association with the chromid? Mrsa smart agar. The customer did not indicate whether the result was reported to the treating physician or impacted treatment decisions. There is no indication or report from the laboratory that the discrepant result led to any adverse event related to any patient's state of health. Though chromid mrsa smart agar (ref. 413050) is not registered, marketed of distributed in the u. S. , a similar product chromid mrsa agar (ref. 43841) is registered in the u. S. Biom? Rieux internal investigation has been initiated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002769706-2017-00302
MDR Report Key6951853
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-10-16
Date of Report2018-05-17
Date Mfgr Received2018-04-25
Device Manufacturer Date2017-07-25
Date Added to Maude2017-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOM
Manufacturer Street5, RUE DES AQUEDUCS
Manufacturer CityCRAPONNE, FR 69290
Manufacturer CountryFR
Manufacturer Postal Code69290
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHROMID? MRSA SMART AGAR
Generic NameCHROMID? MRSA SMART AGAR
Product CodeJSO
Date Received2017-10-16
Catalog Number413050
Lot Number1005860490
Device Expiration Date2017-10-03
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX SA
Manufacturer Address5, RUE DES AQUEDUCS CRAPONNE, FR 69290 FR 69290


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.