TENSOR CP981 FOR MAMMARY MOLE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign report with the FDA on 2017-10-16 for TENSOR CP981 FOR MAMMARY MOLE manufactured by Integra Microfrance S.a.s..

Event Text Entries

[90547825] Integra has completed their internal investigation. Results: evaluation of returned device; the tensor is very hard to use when assembled on 4th position. The axle is distorted and not compliant to our manufacturing specifications. There are numerous impacts on biggest part of tensor , around the 4 drillings. Dhr review; no anomalies that could be associated with the complaint were observed. Complaints history; no adverse trend - first occurrence of this risk for this device. Conclusion: the most probable cause of this distortion is an excessive constraint on the device during use or assembling.
Patient Sequence No: 1, Text Type: N, H10


[90547826] It was reported that the mammary tensor do not tight enough. Medical staff have to call another op on another floor to have another one. Surgery was delayed 45 minutes and haemostasis was difficult as the tensile on the breath was not enough. No patient injured.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2017-00118
MDR Report Key6952468
Report SourceFOREIGN
Date Received2017-10-16
Date of Report2017-09-28
Date of Event2017-09-21
Date Mfgr Received2017-09-28
Device Manufacturer Date2017-06-30
Date Added to Maude2017-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SONIA IRIZARRY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA MICROFRANCE S.A.S.
Manufacturer StreetLE PAVILLON
Manufacturer CitySAINT AUBIN LE MONIAL 03160
Manufacturer CountryFR
Manufacturer Postal Code03160
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTENSOR CP981 FOR MAMMARY MOLE
Generic NamePFM11
Product CodeFTY
Date Received2017-10-16
Returned To Mfg2017-09-28
Catalog NumberCP981
Lot Number2212037
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA MICROFRANCE S.A.S.
Manufacturer AddressLE PAVILLON LE PAVILLON SAINT AUBIN LE MONIAL 03160 FR 03160


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.