MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2017-10-17 for ATLAS CABLE SYSTEM 826-213 manufactured by Warsaw Orthopedics.
[90418575]
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation. Therefore, we are unable to determine the definitive cause of the reported event.
Patient Sequence No: 1, Text Type: N, H10
[90418576]
It was reported that during surgery, the cable broke. The product came in contact with the patient. No fragments remained in the patient. No patient complications were reported as a result of the event.
Patient Sequence No: 1, Text Type: D, B5
[101043141]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[101043142]
The patient underwent hip replacement surgery due to congenital hip dysplasia.
Patient Sequence No: 1, Text Type: D, B5
[115376698]
Additional information: the product was returned with four distinctive failures in the cable. Under microscopic review three of the failures show necking at the ends of the individual strands and a tightening of the original cable twist. These observations are consistent with an overload from tension. The other failure appears to be the result of a cut. There is a uniformity to the angle of the single strands and witness marks can be seen on the face of the strands. If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
[124180337]
If information is provided in the future, a supplemental report will be issued.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1030489-2017-02209 |
MDR Report Key | 6953928 |
Report Source | COMPANY REPRESENTATIVE,FOREIG |
Date Received | 2017-10-17 |
Date of Report | 2018-10-04 |
Date of Event | 2017-09-22 |
Date Mfgr Received | 2018-08-08 |
Device Manufacturer Date | 2016-06-27 |
Date Added to Maude | 2017-10-17 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | STACIE ZIEMBA |
Manufacturer Street | 1800 PYRAMID PLACE |
Manufacturer City | MEMPHIS TN 38132 |
Manufacturer Country | US |
Manufacturer Postal | 38132 |
Manufacturer Phone | 9013963133 |
Manufacturer G1 | WARSAW ORTHOPEDICS |
Manufacturer Street | 2500 SILVEUS CROSSING |
Manufacturer City | WARSAW IN 46582 |
Manufacturer Country | US |
Manufacturer Postal Code | 46582 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ATLAS CABLE SYSTEM |
Generic Name | CABLE |
Product Code | ISN |
Date Received | 2017-10-17 |
Returned To Mfg | 2018-06-18 |
Model Number | NA |
Catalog Number | 826-213 |
Lot Number | 0495809W |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | WARSAW ORTHOPEDICS |
Manufacturer Address | 2500 SILVEUS CROSSING WARSAW IN 46582 US 46582 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-10-17 |