ARCHITECT TOTAL PSA 07K70-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional report with the FDA on 2017-10-17 for ARCHITECT TOTAL PSA 07K70-25 manufactured by Abbott Ireland.

Event Text Entries

[90551010] This report is being filed on an international product, architect total psa, list 7k70, that has a similar product distributed in the us, architect total psa, list number 6c06. An evaluation is in process. A follow-up report will be submitted when the evaluation is complete. Event problem and evaluation codes: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[90551011] The customer reported a patient received a pet scan and radiation therapy due to an elevated architect total psa result. The patient was tested on (b)(6) 2015 on the architect total psa assay as a part of routine follow up for a prostate removal which occurred in 2014. The architect generated a prostate-specific antigen (psa) result of 0. 3 ug/l (sample (b)(6)). The patient was referred to radiation oncology at a private hospital and a pet scan was performed on (b)(6) 2015 which was inconclusive. Due to the inconclusive imaging and the psa result, a course of radiotherapy was commenced at the private hospital on (b)(6) 2015. A baseline psa was taken at the private hospital on the first day of radiotherapy. At a follow up appointment with the radiation oncologist following approximately 5 treatments, the baseline psa result was noted to be <0. 01 ug/l on the roche assay. On (b)(6) the psa was repeated on the roche assay generating a result of <0. 01 ug/l and the unnecessary radiotherapy was discontinued.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3008344661-2017-00091
MDR Report Key6954707
Report SourceFOREIGN,HEALTH PROFESSIONAL
Date Received2017-10-17
Date of Report2017-10-30
Date of Event2017-05-03
Date Mfgr Received2017-10-25
Device Manufacturer Date2014-11-21
Date Added to Maude2017-10-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT IRELAND
Manufacturer StreetDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK
Manufacturer CitySLIGO NA
Manufacturer CountryEI
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT TOTAL PSA
Generic NameTOTAL PSA
Product CodeLTJ
Date Received2017-10-17
Catalog Number07K70-25
Lot Number47139LF00
Device Expiration Date2015-11-14
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT IRELAND
Manufacturer AddressDIAGNOSTICS DIVISION FINISKLIN BUSINESS PARK SLIGO NA EI NA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-10-17