HIGH FLOW INSUFFLATION UNIT UHI-4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-10-18 for HIGH FLOW INSUFFLATION UNIT UHI-4 manufactured by Olympus Medical Systems Corp..

Event Text Entries

[90628999] The referenced uhi-4 was not returned to olympus medical systems corp. (omsc) for evaluation at this time, therefore omsc could not evaluate the uhi-4. The uhi-4 was returned to olympus beijing (osh). Osh evaluated the uhi-4 and found that there was no damage on the exterior appearance of the uhi-4, however the uhi-4 did not start up when the uhi-4 was turned power on. Osh exchanged the main circuit board of the uhi-4 to another main circuit board, the uhi-4 started up without any problem. The referenced uhi-4 is planned to return to omsc. There were no further details provided. The exact cause of the reported event could not be conclusively determined at this time. If additional information, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


[90629000] Olympus was informed that during the preparation of the laparoscopic cholecystectomy, the uhi-4 could not operate. The patient was already under anesthesia. The facility changed the uhi-4 to the other similar device and the procedure was completed. There was no report of the patient? S injury regarding this event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010047-2017-01490
MDR Report Key6957320
Date Received2017-10-18
Date of Report2017-11-22
Date of Event2017-09-19
Date Mfgr Received2017-11-10
Date Added to Maude2017-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR KATSUAKI MORITA
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO-TO
Manufacturer CountryUS
Manufacturer Phone426425177
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHIGH FLOW INSUFFLATION UNIT
Generic NameHIGH FLOW INSUFFLATION UNIT
Product CodeFCX
Date Received2017-10-18
Model NumberUHI-4
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORP.
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO-TO US


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.