ACCURELIEF ACRL-9000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2017-10-18 for ACCURELIEF ACRL-9000 manufactured by Shenzhen Dongdixin Technology Co., Ltd..

Event Text Entries

[89566322] The design history file was examined for the subject device. These devices are designed according to relevant safety standards and passed all compliance tests. The most likely reason that caused burns is that impedance of the electrode used by user is not uniform.
Patient Sequence No: 1, Text Type: N, H10


[89566323] The end-user called reporting a one week-old burn. The end-user underwent a double mastectomy, and decided to have a tummy tuck at the same time. She had a small bulge on her abdomen, and her doctor recommended the tens unit to reduce the scar tissue. The user put the tens unit on for thirty minutes, using separate pads, one on the right side of the belly button, and one on the left - one hour later, she put the pads in the same area, and used the device for another thirty minutes. Pad "a" was ok, but pad "b" caused a second-degree burn the size of a quarter. Pad "b" is reported as having less silicone. The end-user is reported as having sought after/received medical attention for the burns.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005170249-2017-00011
MDR Report Key6957384
Report SourceDISTRIBUTOR
Date Received2017-10-18
Date of Report2017-07-13
Date of Event2017-06-12
Date Mfgr Received2017-07-14
Date Added to Maude2017-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Manufacturer ContactMR KANG JIANPING
Manufacturer StreetBLOCK A, 5TH FLOOR, FUHUA TECH NO. 9116 BEIHUAN ROAD
Manufacturer CitySHENZHEN, GUANGDONG 518108
Manufacturer CountryCH
Manufacturer Postal518108
Manufacturer G1SHENZHEN DONGDIXIN TECHNOLOGY CO., LTD.
Manufacturer StreetNO.3 BUILDING XILIBAIMANG XUSH INDUSTRIAL ESTATE,518108
Manufacturer CitySHENZHEN, GUANGDONG 518108
Manufacturer CountryCH
Manufacturer Postal Code518108
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameACCURELIEF
Generic NameELECTROTHERAPY DEVICE
Product CodeGZJ
Date Received2017-10-18
Model NumberACRL-9000
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSHENZHEN DONGDIXIN TECHNOLOGY CO., LTD.
Manufacturer AddressNO.3 BUILDING XILIBAIMANG XUSH INDUSTRIAL ESTATE,518108 SHENZHEN, GUANGDONG 518108 CH 518108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2017-10-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.