DIMENSION? 10636928

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-19 for DIMENSION? 10636928 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[90648712] Mdr 2517506-2017-00767 was also filed for the same customer inquiry. Quality control was within specification during the test date. Siemens is investigating the incident.
Patient Sequence No: 1, Text Type: N, H10


[90648713] A discordant depressed tacrolimus (tac) result was obtained on a patient sample on the dimension exl system. The result was reported to the physician and questioned. The sample was sent to an alternate lab and tested on high performance liquid chromatography (hplc) and a lower result was obtained. No corrected report was issued. There are no reports of patient intervention or adverse health consequences due to the discordant depressed tac result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00768
MDR Report Key6961154
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-10-19
Date of Report2017-11-15
Date of Event2017-09-26
Date Mfgr Received2017-10-20
Date Added to Maude2017-10-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
Manufacturer StreetREGISTRATION NUMBER 1226181 101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION?
Generic NameDIMENSION? EXL? 200 INTEGRATED CHEMISTRY SYSTEM
Product CodeMLM
Date Received2017-10-19
Catalog Number10636928
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-19