PERCUSSIONAIRE F00012-HT S00009

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-10-19 for PERCUSSIONAIRE F00012-HT S00009 manufactured by Percussionaire Corp..

Event Text Entries

[90753345] The ipv hc is a homecare therapy device for airway clearance. The suspect device was returned to the distributor's biomed technician, who completed the investigation. The device was investigated and functional tested with no issues noted, and subsequently returned into service. During the technician's investigation, no evidence of heat or smoke was found on or in the device. The only anomolies noted were that the gray outer layer of the power cord insulation was shifted, and exposed a bit more of the inner wires' insulation than typical, though the insulation on the inner wires was intact, with no copper wires showing. Also, the fan grills were missing, and the filter cap was missing. This was not a reportable event, but we decided to file this mdr so that our response to the end user's report would be on record.
Patient Sequence No: 1, Text Type: N, H10


[90753346] Home health nurse reported that the device was hot and smoking, and that the wires on the unit's connector were frayed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000524541-2017-00004
MDR Report Key6962415
Date Received2017-10-19
Date of Report2017-10-02
Date of Event2017-07-24
Date Mfgr Received2017-08-16
Device Manufacturer Date2015-05-07
Date Added to Maude2017-10-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSHAWN BURNS
Manufacturer Street1655 GLENGARY BAY RD.
Manufacturer CitySAGLE ID 83860
Manufacturer CountryUS
Manufacturer Postal83860
Manufacturer Phone2082632549
Manufacturer G1PERCUSSIONAIRE CORP.
Manufacturer Street1655 GLENGARY BAY RD.
Manufacturer CitySAGLE ID 83860
Manufacturer CountryUS
Manufacturer Postal Code83860
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePERCUSSIONAIRE
Generic NameIPV HC
Product CodeNHJ
Date Received2017-10-19
Model NumberF00012-HT
Catalog NumberS00009
OperatorHOME HEALTH AIDE
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPERCUSSIONAIRE CORP.
Manufacturer Address1655 GLENGARY BAY RD. SAGLE ID 83860 US 83860


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-19

© 2026 FDA.report