INNOVANCE HEPARIN 10873448

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-20 for INNOVANCE HEPARIN 10873448 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[90675327] Siemens healthcare diagnostics inc. Is investigating the cause of the discordant, low innovance heparin patient result that was generated on the cs-2500 coagulation analyzer.
Patient Sequence No: 1, Text Type: N, H10


[90675328] A discordant, low heparin patient result (0. 47 iu/ml) using innovance heparin kit lot 46475 was generated on the cs-2500 coagulation analyzer compared to the heparin patient result (0. 83 iu/ml) generated with berichrom heparin kit lot 47293. The innovance heparin patient result (0. 47iu/ml) was reported to the physician. The same patient sample was processed at the same time by the both methods on the same cs-2500 coagulation analyzer on the same day. The patient result generated with innovance heparin kit lot 46475 did not match the clinical expectation of the patient. The heparin result expected was ~0. 80 iu/ml. There was no known impact or adverse health consequence to the patient due to the reported low innovance heparin result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00121
MDR Report Key6964424
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-10-20
Date of Report2017-10-30
Date of Event2017-09-19
Date Mfgr Received2017-10-23
Device Manufacturer Date2016-08-04
Date Added to Maude2017-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLORIANN RUSSO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242287
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINNOVANCE HEPARIN
Generic NameINNOVANCE HEPARIN
Product CodeKFF
Date Received2017-10-20
Catalog Number10873448
Lot Number46475
Device Expiration Date2017-12-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, GM 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.