MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-23 for NUCLISENS? EASY MAG? DISPOSABLE VESSEL 280135 manufactured by Biomerieux Sa.
[90756435]
A customer in the united states notified biom? Rieux of a performance issue associated with nuclisens? Easy mag? Disposable vessel ((b)(4)). The customer reported that the tip was missing from the cartridge after run completion on two (2) different easymags (em 5 and em 9). On both occurrences, the issue was on cartridge position a the second tip position. After the runs completed, the second tip from the left on both affected tip cartridges were missing. The customer reported this lead to a delay in reporting results less than 24 hours, and there no indication or report that the discrepant result led to any adverse event related to a patient's state of health. An internal biom? Rieux investigation will be initiated.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3002769706-2017-00308 |
MDR Report Key | 6969078 |
Report Source | HEALTH PROFESSIONAL,USER FACI |
Date Received | 2017-10-23 |
Date of Report | 2018-02-14 |
Date Mfgr Received | 2018-01-18 |
Device Manufacturer Date | 2017-03-03 |
Date Added to Maude | 2017-10-23 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | MS. ELLEN WELTMER |
Manufacturer Street | 595 ANGLUM ROAD |
Manufacturer City | SAINT LOUIS MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer G1 | BIOMERIEUX SA |
Manufacturer Street | 5, RUE DES BERGES CEDEX 01 |
Manufacturer City | GRENOBLE, 38024 |
Manufacturer Country | FR |
Manufacturer Postal Code | 38024 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | NUCLISENS? EASY MAG? DISPOSABLE VESSEL |
Generic Name | NUCLISENS? EASY MAG? DISPOSABLE VESSEL |
Product Code | JJH |
Date Received | 2017-10-23 |
Catalog Number | 280135 |
Lot Number | Z108HB |
Device Expiration Date | 2018-08-28 |
Operator | HEALTH PROFESSIONAL |
Device Availability | * |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | BIOMERIEUX SA |
Manufacturer Address | 5, RUE DES BERGES CEDEX 01 GRENOBLE, 38024 FR 38024 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-10-23 |