ELECSYS FREE PSA IMMUNOASSAY 03289788160

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2017-10-23 for ELECSYS FREE PSA IMMUNOASSAY 03289788160 manufactured by Roche Diagnostics.

Event Text Entries

[90852072] (b)(4). The event occurred in:(b)(6).
Patient Sequence No: 1, Text Type: N, H10


[90852073] The customer experienced an issue where the elecsys free psa immunoassay results were higher than the elecsys total psa immunoassay results for one patient. The customer used a cobas e 411 immunoassay analyzer with a serial number of (b)(4). The initial free psa result was 0. 426 ng/ml and the initial total psa result was 0. 165 ng/ml. The sample was repeated with a free psa result of 0. 427 ng/ml and a total psa result of 0. 151 ng/ml. An aliquot of the same sample was tested in another laboratory using the same elecsys test. The free psa result was 0. 440 ng/ml and a total psa result of 0. 12 ng/ml. The results were not reported outside of the laboratory. There was no allegation of an adverse event. Refer to the medwatch with a1 patient identifier (b)(6) for the total psa assay.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02401
MDR Report Key6969211
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2017-10-23
Date of Report2018-01-03
Date of Event2017-10-03
Date Mfgr Received2017-10-09
Date Added to Maude2017-10-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameELECSYS FREE PSA IMMUNOASSAY
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2017-10-23
Model NumberNA
Catalog Number03289788160
Lot Number20987702
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-23