Q2 EXTENSION SET, 17 INCH 95902

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2017-10-23 for Q2 EXTENSION SET, 17 INCH 95902 manufactured by Quest Medical, Inc..

Event Text Entries

[90771741] At the time of this report, no patient information was provided. The device was discarded by the customer and the lot number was not provided. A review of the device history record could therefore not be conducted. A review of the complaint and non-conformance records for this product code in the last 24 months was conducted and no similar issue was identified. The manufacturing process for this product line was also reviewed and no quality issues were identified. Quest medical has implemented several internal quality controls to ensure only conforming products are shipped out including but not limited to in-process inspection, final inspection and a dhr review prior to product release. A supplemental medwatch will be submitted if additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


[90771742] Report received states that chemo leaked from below filter where the tubing is bonded to the filter. The medication being infused was etoposide. There were no patient complications resulting from the alleged issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1649914-2017-00080
MDR Report Key6970841
Report SourceUSER FACILITY
Date Received2017-10-23
Date of Report2017-10-19
Date Mfgr Received2017-09-21
Date Added to Maude2017-10-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTOSAN ONOSODE
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal75002
Manufacturer Phone9723326338
Manufacturer G1QUEST MEDICAL, INC.
Manufacturer StreetONE ALLENTOWN PARKWAY
Manufacturer CityALLEN TX 75002
Manufacturer CountryUS
Manufacturer Postal Code75002
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameQ2 EXTENSION SET, 17 INCH
Generic NameINTRAVASCULAR ADMINISTRATION SET
Product CodeFPK
Date Received2017-10-23
Model Number95902
Lot NumberNOT PROVIDED
OperatorNURSE
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerQUEST MEDICAL, INC.
Manufacturer AddressONE ALLENTOWN PARKWAY ALLEN TX 75002 US 75002


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2017-10-23