SYMPHION? TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES FG-0202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-10-24 for SYMPHION? TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES FG-0202 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[90779255] The reported lot number could not be matched to the reported device. Therefore, the lot expiration and device manufacture dates are unknown at this time. (b)(4). Manufacturer: smc (b)(4). The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed. If any further relevant information is identified, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[90779256] It was reported to boston scientific corporation on (b)(6) 2017 that a symphion fluid management accessory device was used during a hysteroscopic myomectomy procedure performed on (b)(6) 2017. According to the complainant, during set-up, the controller detected faulty pressure sensor error message. The procedure was completed with another symphion fluid management kit. There were no patient complications as a result of the event. The patient's condition at the conclusion of the procedure was reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2017-03102
MDR Report Key6972516
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-10-24
Date of Report2017-10-04
Date of Event2017-09-29
Date Mfgr Received2017-10-04
Date Added to Maude2017-10-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSYMPHION? TISSUE REMOVAL SYSTEM FLUID MANAGEMENT ACCESSORIES
Generic NameINSUFFLATOR, HYSTEROSCOPIC, FLUID, CLOSED-LOOP RECIRCULATION WITH CUTTER-COAGULA
Product CodePGT
Date Received2017-10-24
Model NumberFG-0202
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.