MAUDE MDR 6976138

MDR report key
6976138
Report number
9612501-2017-06485
Event key
0
Event type
3
Date of event
2017-10-04
Date received
2017-10-25
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
1
Health professional
3
Initial report to FDA
3
Event location
3

Manufacturer Contact#

Contact
SAHRON MURPHY
Address
60 MIDDLETOWN AVE NORTH HAVEN CT 06473 US
Phone
203-203-2034
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1BLUNTPORT PLUSGOUGE, SURGICAL, GENERAL & PLASTIC SURGERYDAVIS & GECK CARIBE LTDGDH179775P179775PJ7A1320XY R

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12017-10-250

Event Narratives#

N

Patient 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

D

Patient 1

ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, PRIOR TO USE THE DEVICE, SURGEON NOTED THAT A RUBBER PART INSIDE BLUNT GRIP WAS DISENGAGED, THEREFORE, THE DEVICE COULD NOT BE USED AT ALL. NOTHING FELL INTO THE PATIENT'S CAVITY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE STATUS OF THE PATIENT IS NO PROBLEM.

N

Patient 1

EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. THE VISUAL INSPECTION OF THE RETURNED PRODUCT NOTED; THE INNER SEAL OF THE THREADED ANCHOR WAS PROTRUDING FROM THE DEVICE. THE OBTURATOR AND TROCAR WERE RECEIVED. THE CANNULA WAS NOT RECEIVED. THE CONDITION OF THE DEVICE PRECLUDES FUNCTIONAL TESTING. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. ANALYSIS CONCLUDED THERE WERE NO ASSEMBLY COMPONENT RELATED FAILURES. REPLICATION OF THE PROTRUDING SEAL MAY OCCUR WHEN MISHANDLED DURING CLINICAL APPLICATION. SHOULD NEW INFORMATION BECOME AVAILABLE, THE FILE WILL BE RE-OPENED AND THE INVESTIGATION SUMMARY WILL BE AMENDED AS APPROPRIATE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.