DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA 314.743

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2006-04-07 for DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA 314.743 manufactured by Synthes (usa).

Event Text Entries

[449672] During an intramedullary rodding of the left tibia, the hex end of the drive shaft broke, causing the reamer head to lose its connection from the drive shaft. The reamer head was lodged in the medullary canal. The reamer head was removed, completely intact, however an x-ray revealed two fragments of the drive shaft remained in the patient. The fragments will not be removed.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1719045-2006-00048
MDR Report Key697919
Report Source05,07
Date Received2006-04-07
Date of Report2006-03-09
Date of Event2006-03-09
Date Mfgr Received2006-03-09
Date Added to Maude2006-04-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactJODI TEMPLE
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES
Manufacturer Street1101 SYNTHES AVENUE
Manufacturer CityMONUMENT CO 80132
Manufacturer CountryUS
Manufacturer Postal Code80132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA
Generic NameRIA DRIVE SHAFT
Product CodeGFC
Date Received2006-04-07
Model NumberNA
Catalog Number314.743
Lot NumberNI
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key687008
ManufacturerSYNTHES (USA)
Manufacturer Address* WEST CHESTER PA 19380 US
Baseline Brand NameDRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA
Baseline Generic NameRIA DRIVE SHAFT
Baseline Model NoNA
Baseline Catalog No314.743
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2006-04-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.