PROVOX VEGA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2017-10-26 for PROVOX VEGA manufactured by Atos Medical Ab.

Event Text Entries

[90320640] Provox vega. Investigation: investigation of product was not possible since no product was returned. The complaint concerns a leakage around the valve, and there is no information about how long the prosthesis has been used. There is no reason to believe that the leakage around prosthesis is due to product fault. Discussion: a leakage around the puncture is most often related to the quality of the puncture. A weak or miss shaped puncture gives less retention force towards the voice prosthesis t-flange and e-flange, this means that a leakage around may occur. If this is the case, a use vega xtraseal or xtraflange might help. It is also possible to use vega xtraseal together with an xtraflange, which might give a better fixation of the device. The ifu describes what to look for if there could be a problem with the puncture a leakage around may also occur if the tracheal flange and/or the esophagus flange are heavily affected by biofilm formation, or if the prosthesis has been damaged during insertion by wrong handling e. G. By use of sharp instrument. An external physician was consulted: it is almost impossible with this information to give a good explanation of the events. The malnutrition / cachexia can have many causes, and pneumonia can easily complicate a case with cachexia, also without aspiration. Periprosthetic leakage can be related to cachexia as well. In general, patients with a laryngectomy do not have a higher risk of pneumonia. In case of leakage, mild aspiration rarely leads to pneumonia. Only chronic aspiration or non-treated aspiration is dangerous. So that is why the surgeon performed the closure. Probably it failed because the patient was cachectic, but many other reasons are possible for failed surgery. If the leakage cannot be stopped by conservative measures, in general a cuffed cannula is needed until the leakage is stopped. I have no information on that. A salvage surgery can also be performed. So, based on the information you give, it is difficult to really explain al events and their relationship conclusion/action: no reason to believe any product failure, no reason for corrective or preventive actions.
Patient Sequence No: 1, Text Type: N, H10


[90320641] This is the information that was received from the atos medical local representative: the patient died of pneumonia on (b)(6). He had been in the hospital since (b)(6). Dr. Said that the patient seemed to lose quite a bit of weight lately. Leakage around the prosthesis started to occur and hadn't improved. The patient became pneumonia due to voice prosthesis. Additional information was received after our sales rep visited the hospital. Our sales rep revisited the hospital to ask the patients the following; the doctor seemed to have second thought for now that the provox vp was not directly-related to the outcome. - do you know if they did anything to stop the leakage around? The doctor tried to stop the leakage through surgery operation to three-layer-close the te puncture site with pectoral major musculocutaneous flap after extracting the vp by referring to the medicine papers below. Once the closure seemed succeeded but the flap was detached to leak again. - were there other comorbidities as a possible cause for the pneumonia? The patient was suffering from spondylitis. He could not walk without walking aid. He seemed so thin that have poor nutrition. - how long was the vp in situ, when did the leaking start (time between start leakage, diagnose leakage, diagnose pneumonia), who and how was it diagnosed? When the patient was admitted, the leakage around had already been occurred. - "patient seemed to lose quite a bit of weight lately": what is lately, what is the relation with the pneumonia, or other diseases? See above. The doctor presumed it was because of his malnutrition and inconvenience from spondylitis. - leakage around can be caused by a too long vp (pistoning) solved by replacement with a shorter device the doctor did not think so. - other (underlying) causes for leakage around are not vp-related, but tep tract-related, such as: o reflux, o recurrence, o diabetes, x malnutrition, o rt/crt, o thyroid dysfunction, o etc. - did they perform assessments to diagnose the leakage and/or the enlargement of the fistula, and was there an adequate treatment? See the first answer. The doctor diagnosed the size of the fistula was too enlarged to cover with xtraflange or xtraseal and operated to close the site.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8032044-2017-00012
MDR Report Key6979357
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2017-10-26
Date of Report2017-10-13
Date of Event2017-09-18
Date Mfgr Received2017-10-13
Date Added to Maude2017-10-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KAROLINA NILSSON
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer PostalSE-24222
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer Postal CodeSE-24222
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROVOX VEGA
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2017-10-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressKRAFTGATAN 8 HORBY, SE-24222 SW SE-24222


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2017-10-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.