BERICHROM ANTITHROMBIN III 10446672

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-10-30 for BERICHROM ANTITHROMBIN III 10446672 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[91227179] Siemens healthcare diagnostics has investigated the provided information to determine the cause of the discordant, low atiii patient sample result on the atellica coagulation 360 system. No product non-conformance could be identified and a possible cause of the event may be related to sample handling. The instrument and reagent are performing according to specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


[91227180] A discordant, low berichrom antithrombin iii (atiii) patient sample result of <15% was generated on the atellica coag 360 system and was not reported to the physician. The same sample was tested again on the same system and an atiii result of 95. 2% was generated. The same patient sample was repeated on a bcs xp system and an atiii result (not provided by customer) was generated that was consistent with the 95. 2% result on the atellica coag 360 system. All quality control results were within range. There are no reports of patient intervention or adverse health consequence due to the discordant, low atiii result.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2017-00119
MDR Report Key6986810
Date Received2017-10-30
Date of Report2017-10-30
Date of Event2017-10-05
Date Mfgr Received2017-10-05
Device Manufacturer Date2017-02-10
Date Added to Maude2017-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactLORIANN RUSSO
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242287
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer StreetEMIL VON BEHRING STRASSE 76
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameBERICHROM ANTITHROMBIN III
Generic NameBERICHROM ANTITHROMBIN III
Product CodeJPE
Date Received2017-10-30
Catalog Number10446672
Lot Number46999
Device Expiration Date2018-11-10
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-30