COBAS 6000 C (501) MODULE C501 04745914001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-30 for COBAS 6000 C (501) MODULE C501 04745914001 manufactured by Roche Diagnostics.

Event Text Entries

[91446201] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[91446202] The customer received questionable high chol2 cholesterol gen. 2 results for 11 patient samples. Data was only provided for four patient samples. Patient 1 initial result was 607 mg/dl and the repeat result was 224 mg/dl. Patient 2 initial result was 417 mg/dl and the repeat result was 117 mg/dl. Patient 3 initial result was 575 mg/dl and the repeat result was 222 mg/dl. Patient 4 initial result was 495 mg/dl and the repeat result was 182 mg/dl. The erroneous results were reported outside the laboratory. The repeat results were believed to be correct. There was no adverse event. The reagent lot number was 26208301 with an expiration date of 31-mar-2018. The field service representative instructed the customer to performed precision testing which was within specifications.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2017-02464
MDR Report Key6988208
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-10-30
Date of Report2017-11-09
Date of Event2017-09-25
Date Mfgr Received2017-10-13
Date Added to Maude2017-10-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1HITACHI HIGH TECH CORP.
Manufacturer Street882 ICHIGE HITACHINAKA NA
Manufacturer CityIBARAKI 312-8504
Manufacturer CountryJA
Manufacturer Postal Code312-8504
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS 6000 C (501) MODULE
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeCHH
Date Received2017-10-30
Model NumberC501
Catalog Number04745914001
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.