[91929474] Sequencing interpretation: the molecular presence of both thymine and guanine at the polymorphic site indicative of the intron 5 silencing site of gypb c. 270+5g>t in the presence of a homozygous s+ allele combination represents an s+ individual1,2. The sample also was polymorphic at gypb c. 270+3g>a which interfered with probe binding. Interfering polymorphisms are listed as limitations in the precisetype hea molecular test package insert.
Patient Sequence No: 1, Text Type: N, H10


[91929475] The customer reported a possible discrepancy. The donor is s- using the bioarray hea molecular beadchip kit; serology results were s+.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005967741-2017-00027
MDR Report Key6990547
Date Received2017-10-31
Date of Report2017-10-31
Date of Event2017-09-27
Date Mfgr Received2017-09-27
Device Manufacturer Date2017-06-07
Date Added to Maude2017-10-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. LUZ VILLAMIZAR
Manufacturer Street35 TECHNOLOGY DRIVE SUITE 100
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084449591
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHEA 1.2 BEADCHIP KIT, SLIDE
Generic NameHEA 1.2 BEADCHIP KIT, SLIDE
Product CodePEP
Date Received2017-10-31
Model Number800-20202-08
Catalog Number800-20202-08
Lot Number17-264-V
Device Expiration Date2018-01-31
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOARRAY SOLUTIONS LTD.
Manufacturer Address35 TECHNOLOGY DRIVE SUITE 100 WARREN NJ 07059 US 07059


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-31