MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2017-10-31 for DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[91210806] The customer contacted the siemens customer care center (ccc) to report the discordant carbon dioxide results. The ccc remotely aligned server 2 probes, performed sodium hydroxide clean on reagent and primed server 2 probes. The customer ran quality control (qc) and precision, which were acceptable. The customer ran precision testing and the results were acceptable. The cause of the discordant carbon dioxide results is unknown. The instrument is performing according to the specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


[91210807] Discordant falsely low, carbon dioxide (co2) results were obtained on patient samples on a dimension vista 1500 instrument. The discordant results were not reported to the physician(s). The samples were repeated on an alternate instrument, resulting higher. The corrected results were reported to the physician(s). There are no reports of patient intervention or adverse health consequences due to the discordant co2 results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2017-00795
MDR Report Key6991617
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2017-10-31
Date of Report2017-10-31
Date of Event2017-10-06
Date Mfgr Received2017-10-06
Device Manufacturer Date2014-03-14
Date Added to Maude2017-10-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARGARITA KARAN
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243105
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Street101 SILVERMINE ROAD REGISTRATION NUMBER: 1226181
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA 1500
Generic NameDIMENSION VISTA 1500
Product CodeCHS
Date Received2017-10-31
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No0
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address500 GBC DRIVE PO BOX 6101 NEWARK DE 197146101 US 197146101


Patients

Patient NumberTreatmentOutcomeDate
10 2017-10-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.