VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-01 for VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[92054617] The investigation determined that lower than expected vitros dgxn results were obtained from a single level of vitros vitros performance verifier (pv) fluid using vitros dgxn slide lot 1913-0246-3005, tested on a vitros 5600 integrated system s/n (b)(4). The investigation was unable to determine an assignable cause of the lower than expected vitros dgxn results. Neither a vitros dgxn nor a vitros 5600 system issue cannot be ruled out as a review of condition codes posted by the system could indicate either a system or reagent issue. Following service by an ortho field engineer, which included adjustments to the immuno wash fluid assembly, successful calibration and acceptable quality control results were obtained with an alternative reagent lot.
Patient Sequence No: 1, Text Type: N, H10


[92054618] A customer obtained lower than expected vitros digoxin (dgxn) results from a vitros performance verifier (pv) fluid. The result was generated using vitros chemistry products dgxn slides processed on a vitros 5600 integrated system. Vitros tdm pv iii lot l53841, vitros dgxn result 1. 86, 1. 93, 1. 95 and 1. 97 ng/ml versus dgxn expected result 2. 56 ng/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action if patient samples were affected. The customer made no allegations that patient sample results were affected. There was no report of patient harm as a result of this event. (b)(4). This report is number 3 of 4 mdr? S for this event. Four (4) 3500a forms are being submitted for this event as 4 devices were involved.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319809-2017-00181
MDR Report Key6992782
Date Received2017-11-01
Date of Report2017-11-01
Date of Event2017-10-03
Date Mfgr Received2017-10-03
Device Manufacturer Date2017-08-01
Date Added to Maude2017-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS DGXN SLIDES
Generic NameIN VITRO DIAGNOSTIC
Product CodeLFM
Date Received2017-11-01
Catalog Number8343386
Lot Number1913-0246-3005
ID Number10758750004782
Device Expiration Date2018-01-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.