VITROS 5,1FS CHEMISTRY SYSTEM 6801890

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2017-11-01 for VITROS 5,1FS CHEMISTRY SYSTEM 6801890 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[92069736] The investigation determined that higher than expected vitros valp quality control results were obtained from vitros and non-vitros qc fluids run on a vitros 5,1fs chemistry system. A definitive assignable cause could not be determined. However, the most likely assignable cause is instrument related. Maintenance actions following monthly maintenance described in the ortho operator? S guide likely resolved the issue. There is no indication a reagent issue contributed to the event. The investigation determined that the likely assignable cause of this event was instrument related.
Patient Sequence No: 1, Text Type: N, H10


[92069737] A customer observed higher and lower than expected vitros valp quality control results from vitros and non-vitros biorad control fluids processed on the vitros 5,1 fs chemistry system. Vitros tdm i lot m5701 vitros results 19. 0 and 18. 52 ug/ml versus expected 29. 2 ug/ml. Vitros tdm ii lot n5702 vitros results 53. 45 and 51. 46 ug/ml versus expected 68. 9 ug/ml. Non-vitros biorad l3 lot 40943 vitros result 39. 68 ug/ml versus expected 29. 0 ug/ml. Biased results of the direction and magnitude observed could lead to inappropriate physician action. Ortho was not made aware of any patient sample results being affected. However, the investigation cannot definitively conclude that patient sample results were not affected or would not be affected if the event were to recur undetected. There was no allegation of patient harm as a result of this event. This report corresponds to ortho clinical diagnostics inc. Complaint numbers (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1319681-2017-00100
MDR Report Key6993502
Date Received2017-11-01
Date of Report2017-11-01
Date of Event2017-09-29
Date Mfgr Received2017-10-04
Device Manufacturer Date2005-07-22
Date Added to Maude2017-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal Code14626
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVITROS 5,1FS CHEMISTRY SYSTEM
Generic NameCHEMISTRY ANALYZER
Product CodeLEG
Date Received2017-11-01
Catalog Number6801890
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-01

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.