STATIC AND DYNAMIC COMPLIANCE 777400-102

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2017-11-01 for STATIC AND DYNAMIC COMPLIANCE 777400-102 manufactured by Vyaire Medical, Inc.

Event Text Entries

[91201666] Vyaire complaint #: (b)(4). Results of investigation: the vyaire factory service (fs) technician received the suspected device and performed a failure investigation. The reported issue was confirmed. The device would not power on. Also, a burnt analyzer board and solenoid cable were found during the investigation. It is unknown what the root cause of the failure is.
Patient Sequence No: 1, Text Type: N, H10


[91201667] The customer reported that the vmax module absolutely failed to power on. There was no patient involvement at the time of the reported issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2017-06905
MDR Report Key6994319
Report SourceDISTRIBUTOR,FOREIGN
Date Received2017-11-01
Date of Report2017-11-01
Date of Event2017-07-17
Date Mfgr Received2017-10-24
Device Manufacturer Date2004-12-01
Date Added to Maude2017-11-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKRISTIN GRAF
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer G1VYAIRE MEDICAL, INC
Manufacturer Street1100 BIRD CENTER DR.
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2017-11-01
Returned To Mfg2017-10-10
Catalog Number777400-102
Device AvailabilityR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerVYAIRE MEDICAL, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2017-11-01

© 2026 FDA.report