MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,foreign report with the FDA on 2017-11-01 for STATIC AND DYNAMIC COMPLIANCE 777400-102 manufactured by Vyaire Medical, Inc.
[91201666]
Vyaire complaint #: (b)(4). Results of investigation: the vyaire factory service (fs) technician received the suspected device and performed a failure investigation. The reported issue was confirmed. The device would not power on. Also, a burnt analyzer board and solenoid cable were found during the investigation. It is unknown what the root cause of the failure is.
Patient Sequence No: 1, Text Type: N, H10
[91201667]
The customer reported that the vmax module absolutely failed to power on. There was no patient involvement at the time of the reported issue.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2021710-2017-06905 |
MDR Report Key | 6994319 |
Report Source | DISTRIBUTOR,FOREIGN |
Date Received | 2017-11-01 |
Date of Report | 2017-11-01 |
Date of Event | 2017-07-17 |
Date Mfgr Received | 2017-10-24 |
Device Manufacturer Date | 2004-12-01 |
Date Added to Maude | 2017-11-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | KRISTIN GRAF |
Manufacturer Street | 22745 SAVI RANCH PARKWAY |
Manufacturer City | YORBA LINDA CA 92887 |
Manufacturer Country | US |
Manufacturer Postal | 92887 |
Manufacturer G1 | VYAIRE MEDICAL, INC |
Manufacturer Street | 1100 BIRD CENTER DR. |
Manufacturer City | PALM SPRINGS CA 92262 |
Manufacturer Country | US |
Manufacturer Postal Code | 92262 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | STATIC AND DYNAMIC COMPLIANCE |
Generic Name | CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION |
Product Code | BTY |
Date Received | 2017-11-01 |
Returned To Mfg | 2017-10-10 |
Catalog Number | 777400-102 |
Device Availability | R |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | VYAIRE MEDICAL, INC |
Manufacturer Address | 22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2017-11-01 |